dc.description.abstract |
Objective: The objective is to develop and validate stability indicating reverse-phase
HPLC method for estimation of Sitagliptin and Simvastatin in bulk and
pharmaceutical dosage form.
Experimental Work: The proposed RP-HPLC method utilizes Gradient separation
mode with Phenomenex C12, (250 mm X 4.6 mm i.d., 4 μm particle size) column, at
ambient temperature, optimum mobile phase consisted of acetonitrile and 50 mM
Potassium dihydrogen phosphate buffer , apparent pH of buffer was adjusted to 6.0
with Potassium Hydroxide solution, effluent flow rate monitored at 2.0 ml/min and
UV detection has been done at wavelength of 218 nm. The method was successfully
validated in accordance to ICH guidelines Q2 (R1) in terms of specificity, linearity,
precision, recovery, sensitivity and robustness. The drug product was exposed to
acidic, alkaline, photolytic and oxidative stress conditions and the stressed samples
were analyzed by the proposed method Results & Discussions: The retention time of sitagliptin was 5.9 min and simvastain
was 14.5. The described method was linear over a range of 20-220 μg/ml for
sitagliptin and 8-88 μg/ml for simvastatin. The correlation coefficient for both drugs
were found to be 0.999. The accuracy of the proposed method were determined by
recovery studies and found to be 99.21% to 99.54% for sitagliptin and 98.79% to
101.2 for simvastatin. The limit of detection for sitagliptin was 4.79 μg/ml and the
limit of quantification is 14.53 μg/ml respectively. The limit of detection for
simvastatin was 1.54 μg/ml and the limit of quantification is 4.66 μg/ml respectively
There is complete separation of degradation and simvastatin and sitagliptin peak,
which demonstrate the specificity of assay method for estimation of sitgliptin and
simvastatin in presence of its degradation products.
Conclusion: All the method was found to be simple, accurate, economical, robust and
reproducible. There was no interference of any degradants and excipients in the
determination of drugs in tablets so it can be employed as a stability‐indicating one.
The method can be successfully applied for routine QC analysis. |
en_US |