Abstract:
A simple, accurate, precise RP- HPLC method and forced degradation was developed
and validated for simultaneous estimation of Amiloride Hydrochloride and
Furosemide in tablet dosage form. The RP-HPLC method has shown adequate
separation for Amiloride hydrochloride and Furosemide from its degradation
products. The separation was achieved on an Phenomenex luna ODS C18 (250 mm ×
4.6 mm i.d., 5 μm particle size) with an isocratic mixture of 0.1 % TEA in Water pH
3.5 with Ortho phosphoric acid: Acetonitrile (50:50 %v/v ). The mobile phase at a
flow rate of 1.0 ml/min, Injection volume 20 μl and wavelength of detection was kept
at 283 nm. The retention time for Amiloride Hydrochloride and Furosemide was
1.810 min and 4.269 min respectively. The linearity of the proposed method was
investigated in the range of 40-60 μg/ml and 320-480 μg/ml for Amiloride
hydrochloride and Furosemide, respectively. Correlation coefficient was 0.999 and
0.998 for Amiloride hydrochloride and Furosemide respectively. Forced degradation
study was carried out on combined dosage form as per ICH guideline and it was exposed to hydrolysis (acid and base hydrolysis), oxidative and thermal conditions to
apply stress. Proposed method was validated as per ICH guidelines for linearity,
accuracy, precision and robustness for estimation of Amiloride hydrochloride and
Furosemide in commercially available pharmaceutical dosage form and results were
found to be satisfactory. The developed and validated RP-HPLC method can be used
successfully for marketed formulations.